APPENDIX
Table 4: Acceleration and Deceleration Times
Rotor Number
220.95 V09
221.54 V04
221.55 V04
221.57 V02
221.59 V06
221.60 V03
221.75 V03
221.77 V01
221.78 V01
221.79 V01
221.80 V01
Table 5: Error Messages
Error-No.
Describtion
No.Nr.:
1
Imbalance
2
Imbalance sensor is defective
14
Problem with speed sensor
15
Standstill monitoring defective
16
Wrong dircention of rotoration of the motor
18
Device gets too warm, cool down
33
Open lid while motor is running
34
Lid contact defective
43
Undervoltage frequency converter
44
Overvoltage frequency converter
47
Error current detection
55
Overspeed
80
Memory Error intern EEPROM
81
Memory Error intern EEPROM, program data
99
Not permitted
© Hermle Labortechnik GmbH
Acceleration Values
Level 0
Level 9
140
19
295
34
293
34
198
22
198
23
319
30
198
18
182
23
241
26
240
25
242
26
Acceleration Time
from 0 rpm-> n
max
Z287A_V1.19
_eng
Deceleration Values
Level 0
145
317
317
212
211
334
213
193
250
253
252
in seconds
Deceleration Time
von n
max
Level 9
21
38
38
23
24
33
27
29
31
32
33
-> 0 rpm