For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 446 K, is a refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. We offer the 120 V voltage centrifuge as a 15 A and 20 A version.
Instructions for Disposal (see P. 34) Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen In-Vitro-Diagnostic Medical Device TYPE: Type Designation of the Product REF: Order No.
EN 61010-2-020. • The HERMLE Z 446 K is "non-explosion proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
Do not run the centrifuge, if mechanical or electrical assembly groups have been tampered with, by unauthorized personnel. • Do not use accessories such as rotors and buckets, that are not approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes, made of glass or plastic.
1.7 Installation of the Centrifuge 1.7.1 Unpacking the Centrifuge Model Z 446 K is supplied in a carton. Remove the strap retainer, open the carton, and remove the padding. Lift the centrifuge on both sides (seeFigure 1) with an appropriate number of helpers and place it on the laboratory table.
The bolts of the rotor must be greased with the HERMLE Rotorgrease (Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the drillings or tube racks. The weight difference of the loaded buckets should not exceed 1.0 g.
"TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol- tight". ATTENTION: Autoclaving, mechanical stresses and contamination, by chemicals or other aggressive solvents, can impair the aerosol-tightness of the rotors and buckets.
For all devices delivered before 13.08.2005, the last user is responsible for proper disposal. RoHS Declaration of Conformity HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, Germany, hereby declares that all components produced are in conformity with Directive 2015/863/EU of the European Parliament and of the Council of 31.3.2015 on the restriction of the use of certain hazardous substances in...
Z 326; Z 327; Z 366; Z 446; Z 207 MK; Z 216 MK; Z 32 HK; Z 326 K; Z 327 K; Z 366 K; Z 36 HK; Z 446 K; ZK 496 Einschlägige EG-Richtlinien / Normen / Relevant EC Directives / Standards (EU) 2017/746;...
Page 43
APPENDIX Table 1: Technical Data HERMLE Labortechnik GmbH, 78564 Wehingen Manufacturer Z 446 K Type Dimensions 73 cm Width 67 cm Depth Height 39 cm 111 kg Weight without rotor Max. Speed 16000 min 4 x 750 ml Max. Volume 26331 x g Max.