2.5.2 Recall of Stored Programs

To recall stored programs, press the key "prog" (11), with the lid already closed. Inside the display
"time" (A-3), appears "programm --"(M10). With the adjustable knob (1), pre-select the desired
program number. In the respective displays, the stored values, for that program, will appear.
For all number marked text, please refer to Figure23.
Figure23

2.5.3 Leaving Program Mode

To leave the program mode, press the key, "prog" (11). Then, inside the display "time", appears the
word "programm".
Set the display to "programm--" (M10) with the adjustable knob (1).
For all number marked text, please refer to Figure23.

2.6 Starting and Stopping the Centrifuge

2.6.1 Starting the Centrifuge

Start the centrifuge with either the "start" key (9), or the "quick" key (8). With the "start" key (9),
stored runs or runs with manually pre-selected parameters can be started. When the respective pre-
selected running time has ended, the centrifuge will stop automatically. With the "quick" key (8), start
runs, which will last a few seconds, can be initiated.
By pressing the "quick" key (8), the centrifuge accelerates up to the pre-selected revolution. In the
display "time" (A-3), the passed running time is indicated from the moment the "quick" key (8) is
pressed.By releasing the "quick" key (8), the centrifuge stops and the running time is indicated, until
the lid is opened.
The "quick" key (8) doesn´t work during a stored program.