Maintenance Summary - Care activities
AM110
- Check the window pane at the level sensor of the vacuum rotation filter
Orientation
Procedure
Maintenance Instructions
MS32-6.3
DIM105EN - 26.04.2022
Check the window pane at the level sensor of the vacuum rotation filter. This sensor
is located at the outside of the vacuum rotation filter (see Figure 1). The sensor
monitors the fill level in the cooling lubricant tank. A very dirty window pane may
cause malfunction. Therefore, regular visual inspection from the outside is
necessary. If the Window pane is very dirty, it must be cleaned on the inside.
1.
Figure 1: Location of the level sensor on the vacuum rotation filter
X
Inspection window
Y
Cover plate
Z
Cover of the sensor
Check the window pane at the level sensor of the vacuum rotation filter.
2. Unscrew the cover (Z).
3. Check the window pane on the sensor.
Alternatively, the following procedure can be used!
If the window pane is very dirty, the pane must be cleaned on the inside of the
cooling lubricant tank.