9. Troubleshooting
ID
Message
32
Calibration Data Error ME
33
Line Overvoltage
34
Line Undervoltage
36
Communication Failure
PU
37
Communication Failure
ME
38
Communication Failure SC System
40
Polyphase Connection
Timeout
41
Mismatched Firmware
Versions
43
Mismatched Pw Unit
Versions
44
Unsupported Pw Unit
Version
45
Mismatched Cabinet
Connect
46
Inhibited Cabinet Connect
47
No Available Power Unit
48
Output Overvoltage
220
Table 9-2 Error Message List (Continued)
Protection
type
Message
display only
System
Lock
System
Lock
System
Lock
System
Lock
Lock
System
Lock
System
Lock
System
Lock
System
Lock
System
Lock
System
Lock
System
Lock
Output Off
Cause or Necessary action
The adjustment data is abnormal. Cycle the
power. If it still occurs, contact the seller
(NF Corporation or our agent) from whom
you purchased the product.
The power input voltage is excessive. Check
if the power input is appropriate.
The power input voltage is insufficient.
Check if the power input is appropriate.
There is a communication error between the
firmware and the power unit. Cycle the
power. If it still occurs, contact the seller
(NF Corporation or our agent) from whom
you purchased the product.
There is a communication error between the
firmware and the output measurement
section. Cycle the power. If it still occurs,
contact the seller (NF Corporation or our
agent) from whom you purchased the
product.
The polyphase system communication is
abnormal. Make sure the System Cable is
connected properly.
In the polyphase system, the connection
destination cannot be recognized in the
specified time. Make sure the System Cable
is connected properly, then turn on the
power of all the cabinets in 20 seconds.
In the polyphase system, the main firmware
versions of the cabinets do not match.
All power unit versions do not match.
A power unit of unsupported version is
connected.
In the polyphase system, the output
capacities of the phases do not match.
A cabinet to which the polyphase system
connection is inhibited is connected with the
System Cable.
There is no power unit that can be started.
The output voltage is excessive. This may
occur at an abrupt change in the output
current due to the inductive load.
Also when the remote sensing function set
to FB, check the feedback cable connects
correctly. (See 8.2.2)
PROGRAMMABLE AC POWER SOURCE