4. Advanced Operation
 Turning off the Autocal function
1. Open the Autocal set window. Following two methods are available.
(a) Shortcut operation:
The window is toggled between the AGC setting window and the Autocal setting window
every time you press the keys.
(b) Push the [Misc] soft-key and select [1: AGC/Auto Cal].
2. In the [Auto Cal] item, select [1: OFF]. When [OFF] is selected here, the Autocal calibration
finishes, and the calibration factor is cleared.
3. Use the [Rmt Sense] item to set the Remote sensing function to on, off, or FB, if necessary.
4. Close the window.
156
+