4. After the parameter settings are complete, press [Enter].
5.6 Set the Power-on State (PowerOn)
This menu item is used to control the settings and output state when the power
supply is powered on.
The detailed power-on parameters and output state are as follows.
•
•
•
The procedures to set the menu item are as follows.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
2. Turn the knob to select the PowerOn and press [Enter].
3. Press the Left / Right key or turn the knob to adjust the value of this
•
On: Default value, indicates the beeper is on.
•
Off: Indicates the beeper is off.
At this point, the beeper status takes effect immediately.
The voltage, current, power setting values displayed on the main interface
when the instrument is powered on.
Parameter setting values in the Config menu.
The output state of the power supply, i.e., the [On/Off] key state.
ter the system menu.
parameter.
•
Reset: Default value, indicates when the instrument is powered on, the
instrument will initialize some parameter settings or state.
The affected parameters and the reset information are as shown in the
following table.
Table 5–1 Initial value of the parameter
Parameter
Category
Voltage setting: Vset
Main
interface
Current setting: Iset
Upper limit of voltage: Vlim
Copyright © Itech Electronic Co., Ltd.
Basic Operation
Initial Value
0.002V(10V models)
0.2V(Other models)
One percent of the
rated current of the
instrument
Upper limit value:
0.002V(10V models)
71