4. Select CC and press [Enter] to confirm.
5. Set the total number of steps in the List program and press [Enter] to
6. Set the current, slope and time width of step 1 in turn, and press [Enter] to
7. Set the parameters of step 2 in the same method.
8. Set the number of list repetitions Repeat, and press [Enter] to confirm.
9. Set the End State as Normal, and press [Enter] to confirm.
10. Depending on the test requirement, choose whether to turn on the function
11. Set the currently edited List program name.
Import List Program
If you need to run the List program on the external USB flash drive, you need to
import the List program from the USB flash drive into the instrument.
The procedures are as follows:
1. Insert the USB flash drive into the front panel USB connector.
2. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
3. Select 1. LIST: Off and press [Enter].
4. Press the Left/Right key to select Open and press [Enter].
5. Select USB and press [Enter].
confirm.
confirm.
switch that triggers synchronization.
If you choose not to save, press [Esc] to exit the editing interface.
The format of the List file in the USB flash drive must be in .csv format and
saved in the root directory of the USB flash drive. You can refer to
Program
to customize editing according to the List file template exported to
the USB flash drive.
ter the function menu.
Rotate the knob, and the system will automatically read all .csv format files
in the root directory of the USB flash drive at this time. The interface is dis-
played as follows:
XXX.csv
Not-Load
Load
XXX indicates the name of the List file; YY indicates the serial number of the
current List file; ZZ indicates the total number of the List file.
Copyright © Itech Electronic Co., Ltd.
YY/ZZ
Power Supply Function
Export List