Set Under-Current Protection (Ucp) - ITech IT-M3900D Series User Manual

High power programmable dc power supply
Hide thumbs Also See for IT-M3900D Series:
Table of Contents

Advertisement

Possible Cause
Many reasons can cause OPP, the details are as follows:
How to Set
The operation steps to set OPP are as follows.
1. Press the composite keys [Protect] on the front panel to enter the protection
2. Press the composite keys [Shift]+[Recall] (Protect) on the front panel to en-
3. Rotate the knob to select 3. OPP (Off) and press [Enter].
4. Press the left/right key or rotate the knob to select On and press [Enter] to
5. Set the protection limit Level and the delay time Delay in sequence, and

4.5.4 Set Under-Current Protection (UCP)

Users can enable the UCP function and set the instrument warm-up time
Warm-up, protection limit Level and protection delay time Delay. When the cur-
rent (i.e., the Meter value) is lower than this protection limit and the warm-up
time, delay time are exceeded, the power supply will enter the UCP state.
Possible Cause
Many reasons can cause UCP, the details are as follows:
How to Set
The operation steps to set UCP are as follows.
The set protection limit Level is lower than the power Meter value.
The power supply outputs a high power due to a fault.
menu. (1U Model)
ter the protection menu. (2U Model)
enter the setting interface.
press [Enter] to confirm.
The VFD screen returns to the Protect menu and the OPP setting (take
0.15KW, 1S as an example) is displayed as follows:
PROTECT
3.OPP
0.15KW, 1.000S
The set protection limit Level is greater than the current Meter value.
The external (AC input) inputs a lower current.
The power supply outputs a low current due to a fault.