Adjust Span Coefficients
Thermo Scientific
The Adjust Span Coefficients screens are used to manually adjust the span
coefficients.
The following screen is shown in single range mode and dual or auto range
mode. In dual or auto range modes, "High Range" or "Low Range"
buttons are displayed to indicate the calibration of the high or low
coefficient. The Adjust High Span Coefficient and Adjust Low Span
Coefficient screens function the same way as the (single range) Adjust Span
Coefficient screen.
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Span Coefficients (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Span Coefficients (dual range mode)
The Adjust Span Coefficients screen contains the following information:
Calibrate SO
Coefficient: Adjusts the SO
G
2
range mode.
Calibrate SO
High Range Coefficient: Adjusts the SO
G
2
coefficient when in dual or auto range mode.
coefficient when in single
2
high range
2
450iQ Instruction Manual
Operation
Calibration
3-25