Thermo Scientific 450iQ Instruction Manual page 259

Hydrogen sulfide and sulfur dioxide analyzer
Table of Contents

Advertisement

Thermo Scientific
Optional Permeation Oven
G
The firmware contains the active controls for their application, as well as
self-identification and configuration for "plug and play" style operation.
Each are associated with specific registers of two types:
Modbus registers that are communicated from each microprocessor to
G
the System Controller Board (SCB) via internal RS-485
SNMP registers that are maintained in the software and SCB for health
G
and data processing computation
The Modbus communication system operates on 1 second intervals.
Within those intervals, data treatment like integration (whether analog or
digital) and servo control, are embedded in the module firmware. The SCB
receives the 1 second updates for higher level "software" processing and
control via SNMP registers, some of which is interfaced with the front
panel Graphical User Interface (GUI).
In addition to the operating registers, the 450iQ stores a historical data log
in a MySQL database. The memory is provided on the same micro SD card
where the operating software resides, for which there is capability to store
up to a year of data at 1 minute intervals. Chapter 3, "Operation" describes
how this database is accessed and used including external memory
downloads.
System Description
Firmware
8-9