SO
Bench Settings
2
Thermo Scientific
The SO
Bench Settings screen allows the user to calibrate the SO
2
The calibration in this screen should be performed if the part associated
with the optical bench has been changed/repaired or if the span coefficient
is outside the limit of 0.5 to 2.0.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To Detector Gain Settings
Screen
The Detector Gain Settings screen contains the following information:
Across:
G
Description: Define the actions the user can do.
G
Detector Gain: User can manually set or automatically calibrate the
G
detector gain.
Operation
Settings
bench.
2
3-83
450iQ Instruction Manual