Thermo Scientific 450iQ Instruction Manual page 109

Hydrogen sulfide and sulfur dioxide analyzer
Table of Contents

Advertisement

SO
Bench Settings
2
Thermo Scientific
The SO
Bench Settings screen allows the user to calibrate the SO
2
The calibration in this screen should be performed if the part associated
with the optical bench has been changed/repaired or if the span coefficient
is outside the limit of 0.5 to 2.0.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To Detector Gain Settings
Screen
The Detector Gain Settings screen contains the following information:
Across:
G
Description: Define the actions the user can do.
G
Detector Gain: User can manually set or automatically calibrate the
G
detector gain.
Operation
Settings
bench.
2
3-83
450iQ Instruction Manual