Thermo Scientific 450iQ Instruction Manual page 102

Hydrogen sulfide and sulfur dioxide analyzer
Table of Contents

Advertisement

range is active, the range mode selection status output is at 0 volts.
When the high range is active, the range mode selection status output is
at half of full-scale.
When the high range is active, the concentration must drop to 95% of
the low SO
range for the low range to become active.
2
In addition, each SO
range and analog output has a span coefficient so
2
that each range can be calibrated separately. This is especially necessary
if the two ranges are not close to one another. For example, the low
SO
range is set to 0–50 ppb and the high SO
range is set to 0–10000
2
2
ppb.
450iQ Instruction Manual
Thermo Scientific