Thermo Scientific 450iQ Instruction Manual page 232

Hydrogen sulfide and sulfur dioxide analyzer
Table of Contents

Advertisement

Photo Lamp
Use the following procedure to remove and replace the photo lamp
detector board.
Detector Board
Replacement
Equipment required:
Phillips drive, #1 and #2
1. Unplug cable.
2. Using #2 Phillips drive, unfasten three #6-32 pan head screws.
Figure 7–42. Replacing the Photo Lamp Detector Board pt 1
3. Pull detector cover outwards.
4. Using #1 Phillips drive, unfasten four #4-40 pan head screws.
5. Pull photo lamp detector board outwards.
6. Replace lamp detector board and assemble in reverse order.