Thermo Scientific 450iQ Instruction Manual page 72

Hydrogen sulfide and sulfur dioxide analyzer
Table of Contents

Advertisement

Operation
Settings
3-46
450iQ Instruction Manual
SO
Bench
The SO
2
module. If an item being monitored goes outside the lower or higher alarm
limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>SO
The SO
Across:
G
G
G
G
G
G
Down:
G
G
G
G
G
G
Bench screen provides status and alarms related to the SO
2
Bench screen contains the following information:
2
SO
Bench: This column lists items associated with the SO
2
Value: Displays the current value for each item.
Low Alarm: Displays low alarm for each item.
High Alarm: Displays high alarm for each item.
Units: Displays units for each item.
SO
Bench Temperature: Displays the current bench temperature.
2
User can adjust low and high alarm limits.
Lamp Intensity: Displays the current lamp intensity reading. User
can adjust low and high alarm limits.
Lamp Voltage: Displays the current lamp voltage. User can adjust
low and high alarm limits.
PMT Voltage: Displays the current PMT supply voltage.
PMT Voltage Too Low: Displays OK/Fail for PMT voltage too low.
bench
2
Bench
2
bench.
2
Thermo Scientific