Problem
Alarm – 5V/24V Step Board
Alarm – Instrument
Temperature
Alarm – Analog I/O
Alarm – Digital I/O
Alarm – Auto Bkg
Cal/Check
Alarm – Auto Span
Cal/Check
Possible Cause
Cable connection
Check the cable connections to
that Step POL board.
Fan failure
Replace fan if not operating
properly.
Dirty fan filter
Clean or replace filter.
Overheating PCB
Locate defective PCB reporting
the error and replace if needed.
Defective PCB
Replace Analog board.
Defective PCB
Replace Digital board.
Incorrect high alarm limit
Verify the high limit is correct via
Settings>Status and
Alarms>Concentrations screen.
Instrument background
Recalibrate the instrument.
calibration failed
Incorrect high alarm limit
Verify the high limit is correct via
Settings>Status and
Alarms>Concentrations screen.
Instrument span
Recalibrate the instrument.
calibration failed
Troubleshooting
Troubleshooting Guide