Smart Echo Level Test - Emerson Rosemount 5408 Reference Manual

Level transmitters
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Operation
March 2023
6.5 

Smart echo level test

The function allows you to test the behavior of the transmitter in a real tank environment
without raising the level. During the test, a virtual surface echo is superimposed onto the
radar signal, and the transmitter will output a level corresponding to the echo position.
The test verifies the integrity of the signal processing, and can be used to test the alarm
limits in the host system, output of the transmitter, and transmitter configuration (for
example the upper/lower range values).
Related information
Rosemount 5408:SIS Safety Manual
6.5.1 
Configure the smart echo level
Prerequisites
This test is available for transmitters with:
Device software revision 1.C0 or later
Option code ET
Rosemount Radar Master Plus
Procedure
1. Select Service Tools → Proof Test.
2. Select Configure Smart Echo.
3. In the Smart Echo Level box, enter the desired value.
4. Select Save.
6.5.2 
Perform a smart echo level test
During the test, a virtual surface echo is superimposed onto the radar signal, and the
transmitter will output a level corresponding to the echo position.
Prerequisites
This test is available for transmitters with:
Device software revision 1.C0 or later
Option code ET
Rosemount Radar Master Plus
Prior to and during the test, ensure that:
The product surface is calm.
The smart echo level is not closer than 1 ft. (0.3 m) to the liquid surface (3.3 ft. [1 m] for
the extended range option and 5 ft. [1.5 m] when measuring solids).
The tank is not being filled or emptied.