9. Tighten the set screw.
10. Place a suitable gasket on the tank flange.
11. Lower the antenna assembly into the tank.
12. Tighten the bolts and nuts with sufficient torque for the flange and gasket choice.
Postrequisites
1. Adjust the inclination of the antenna.
2. Connect the air purging system.
Rosemount 5408 and 5408:SIS Level Transmitters
Torque 5 in-lb (0.5 N-m)
H2 mm
Mechanical installation
March 2023
57