Service and troubleshooting
March 2023
Geometry configuration error
Cause
The configured tank geometry results in a too large level measuring range for this device.
Recommended actions
Related information
Reference height
Restore to default settings
Configure transmitter using guided setup
Primary variable configuration error
Cause
The Primary Variable selection is not supported.
Note
Rosemount 5408:SIS only supports level or distance as Primary Variable.
Recommended actions
Related information
Primary variable
Measurement correction configuration error
Cause
The factory measurement correction data is invalid.
Recommended actions
Related information
Restore to default settings
Configure transmitter using guided setup
Threshold configuration error
Cause
The surface threshold configuration is incorrect.
Recommended actions in UIP
96
1. Check tank geometry configuration and reduce Reference Height.
2. If condition persists, restore default settings and reconfigure the device.
1. Change Primary Variable to variable supported by device.
2. Consider purchasing an upgrade of the device to access additional variables.
1. Restore default settings and reconfigure the device.
2. If the condition persists, replace the device.
1. Adjust the threshold as necessary.
2. If condition persists, restore default settings and reconfigure device.