Reference Manual
00809-0100-4408
5.5.3
Radar Master error message
If the
revision on your device requires an update of the Rosemount Radar Master Plus FDI
Package.
5.6
Confirm HART
If using HART based control or asset management systems, confirm the HART capability
of those systems prior to transmitter installation. Not all systems are capable of
communicating with HART Revision 7 protocol. This transmitter can be configured for
either HART Revision 6 or Revision 7.
5.6.1
Switch HART
If the HART configuration tool is not capable of communicating with HART Revision 7, the
device will load a generic menu with limited capability. To switch the HART revision mode
from the generic menu:
Procedure
1. Locate the "Message" field.
2. In the Message field, enter HART6 or HART7 and then 27 trailing spaces.
5.7
Configure transmitter using guided setup
The options available in the Guided Setup wizard include all items required for basic
operation.
Procedure
1. If using an FDI compliant software, then select Overview → Rosemount Radar
2. Select Configure → Guided Setup and follow the on-screen instructions.
Related information
Save a backup file of the device configuration
Rosemount 5408 and 5408:SIS Level Transmitters
Can't initialize Radar Master
revision capability
®
revision mode
®
Master Plus.
error message occurs, then the software
Configuration
March 2023
79