Note Regarding
Calibration
General Requirements
and Recommendations
Thermo Scientific Smart-Vue
Our calibration is carried out according to COFRAC guidelines and are not
COFRAC-certified calibrations. Our calibration tools are verified on a
regular basis by L.N.E., a COFRAC-certified laboratory. Please contact us
for information on calibration carried out by the COFRAC Laboratory.
The machine hosting Smart-Vue Server software runs continuously,
24/7/365, and should be connected to an Uninterruptible Power Supply
(UPS) that protects against power surges and provides power to back up
the machine and components. Refer to Section 21.2 for recommended
machines based on your individual installations. For assistance in
determining the correct machine, please contact Thermo Scientific.
All Smart-Vue products using an AC adapter should also be plugged
into the UPS.
The UPS is also recommended as a power backup for your
communication/network systems.
Weekly manual system testing (e.g., disconnect a sensor from its
module, verify the communications system is working) should be
performed as defined in your Standard Operating Procedure (SOP).
Recommended maintenance and calibration procedures should be
followed.
If you are storing cold products, Thermo Scientific recommends use of a
back-up cooling system (e.g., CO
temperature below the critical level should a power failure occur.
Contact your local sales representative for more information.
Thermo
Qualifications (IQ/OQ) be performed before initial use.
For timely notification of an emergency/alarm, it is critical to establish
primary, secondary and tertiary call-out procedures with escalation so
contact is not dependent on reaching one individual. Ideally, ultimate
contact is with a security department or company with 24/7
monitoring.
Your Smart-Vue Server computer and Smart-Vue Client computers
must use the same Windows Regional Settings.
CAUTION: Modules shall not be placed in environmental conditions
beyond recommended specifications.
®
Wireless Monitoring Solution
2
Scientific
recommends
or LN
) to maintain freezer chamber
2
Installation
and
Administrator Guide v 2.2
Notices
Operational
5