Operation - Ma Measurements; Operation - Switch Test - GE Druck DPI 821 User Manual

Thermocouple calibrator and thermocouple loop calibrator
Table of Contents

Advertisement

Operation - mA measurements

DPI 822 only. To measure a current:
1. Connect the instrument (Figure 5) and, if necessary,
adjust the Set Up (Table 1).
2. Select the applicable mA input task from Select Task
(Table 2/3) and, if necessary, adjust the Settings
(Table 4).
Maximum: 30V
a) with external loop
Figure 5: Example configuration - mA measurement

Operation - Switch test

DPI 822 only. To do tests on a thermocouple switch:
1. Connect the instrument (Figure 6) and, if necessary,
adjust the Set Up (Table 1).
2. Select the applicable switch test from Select Task
(Table 2/3) and, if necessary, adjust the Settings
(Table 5). The display shows the switch condition (open
or closed) in the top right-hand corner.
3. Supply the output values to the system (Table 7).
• Example - "Nudge" output.
a. Use Edit (
) to set a value less than the switch
value.
b. Use the
buttons to change the value in small
increments.
8 - [EN] English
Settings
b) with internal loop
power
power
• Example - "Ramp" output.
a. Set "High" and "Low" values that are applicable to
the switch value (Table 6). Then, to get an accurate
switch value, set a long "Travel" period.
b. Use Start/Stop (
cycle.
4. If necessary, supply the output values in the opposite
direction until the switch changes condition again.
The display shows the applicable values to open and
close the switch.
5. To do the test again, press ESC to reset the values.
Settings
Figure 6: Example configuration - Switch test
) to start and stop the "Ramp"
Edit
Settings
Start
a) Temperature
b) mV
Settings
K346 Issue 2