Installing The Multi-Detector Drip Tray - Waters ACQUITY UPC2 Overview And Maintenance Manual

Photodiode array detector
Hide thumbs Also See for ACQUITY UPC2:
Table of Contents

Advertisement

6.
Remove the protective cover from the PEEK cell inlet tubing, and
connect the tubing to the flow cell inlet, confirming that the label on the
tubing matches the type of detector and flow cell in your system.
Note:
in two lengths:
20.5" long, for dual stack configurations
16" long, for single stack configurations
Use the appropriate length of tubing for your configuration.
7.
Connect the outlet tubing from the outlet of the flow cell to the
convergence manager. Finger-tighten, then use the right-angle tool (nut
extender, p/n 410001941 in the start up kit) to tighten 1/4 turn.

Installing the multi-detector drip tray

If your ACQUITY UPC
components (such as dual CM-Aux modules plus a CM-A module), you must
install one or more multi-detector drip trays. If you have two detectors, the
trays should be positioned under each detector. If you have an additional
module, the tray should be positioned under the detector that sits atop the
module.
Required materials
Multi-detector drip tray kit
To install the drip tray:
1.
Turn the ACQUITY UPC
2.
Snap the extended plastic feet onto the bottom of the detector, and then
snap the anti-skid pads onto the extended plastic feet.
3.
Secure the drip tray to the bottom of the detector, using the screws and
plastic rivets provided in the multi-detector drip tray kit.
a.
Remove the two screws, move the drip tray into place, and then
reinstall the screws to secure the tray.
b.
Install three plastic rivets to further secure the tray.
24
Flow cell tubing for the ACQUITY UPC
2
system has more than one detector or additional
2
PDA detector, laying it on its left-hand side.
2
PDA detector is provided