TCC-100 Operating Instructions
Commands/
Properties
in Chromeleon
TCC_RELAY_2
TempCtrl
Temperature
Vendor
For information about individual commands and parameters and about the command syntax,
open a Chromeleon control panel. Select Command... on the Control menu. The
Commands dialog box is opened. You can also open this dialog box by pressing the F8 key.
In the left list box, click the "+" character beside ColumnOven to display the items
underneath.
For more information (also about any parameter or command that might not be listed here), refer
to the Chromeleon online Help.
Important: If you press the Standby key on the instrument's front panel while the
column compartment is under Remote control, all instruments connected
to a USB hub on the instrument's rear panel are also disconnected from
Chromeleon.
Please note: When the column compartment is connected to data system, the front
panel Remote LED is lighted and column compartment is completely
controlled and programmed via Chromeleon. It is not possible to change
the settings on the instrument's front panel.
While the column compartment is under Chromeleon control, you can
nevertheless display the temperature set point and the leak detection mode
on the instrument's front panel by pressing the Plus or Minus key (set
point) or by pressing both keys simultaneously (leak detection mode).
There are five leak detection modes:
Ld0 = The sensors are turned off.
Ld1 = The sensors are turned on. In Chromeleon, the sensitivity for the gas
Ld2 = The sensors are turned on. In Chromeleon, the sensitivity for the gas
Ld3 = The sensors are turned on. In Chromeleon, the sensitivity for the gas
Description
Relay 2. Click the "+ "character beside the relay name to display the items
underneath:
State−indicates the state of the relay.
Duration−when set, the relay toggles after the specified time.
On−turns the relay on.
Off−turns the relay off.
Turns temperature control on or off.
Range: 5.0...85°C. Click the "+ "character beside the name to display the
items underneath:
Value−indicates the actual temperature.
Nominal−allows you to enter the temperature set point.
UpperLimit−allows you to enter the upper temperature limit.
LowerLimit−allows you to enter the lower temperature limit.
If the temperature set point is outside these values, an error message
appears in the Audit Trail.
Indicates the device manufacturer (read-only).
and humidity sensors has been set to Low.
and humidity sensors has been set to Standard.
and humidity sensors has been set to High.