Thermo Scientific Dionex TCC-100 Operating Instructions Manual page 39

Thermostatted hplc column compartment
Table of Contents

Advertisement

Example for a valid configuration:
Column compartment: IP address
Server:
To configure several TCP/IP-enabled Summit HPLC devices, make sure that every device has
its own IP address. In this example, the following entries would be possible:
2. TCP/IP-enabled device: IP address
3. TCP/IP-enabled device: IP address
Important: Every device on the same network must have a unique IP address.
Important: Do not use 0 or 255 as the last numbers in IP addresses.
• The Gateway Address field must not be empty, although no gateway is available for the
instrument LAN. Therefore, enter the IP address of the device to be configured in this field to
prevent a gateway from being used.
• When you finish entering IP properties, assign them to the module by clicking Assign. After
about 1 second, the instrument's new parameters will be displayed in the Module Information
window.
• Click the Query button to check the settings.
• Click Exit to close the program.
Important: Verify that a unique IP address has been assigned to all Summit HPLC
devices that are connected to the Chromeleon server PC.
For more information, please refer to the CmIPUtil online Help.
192.168.60.51
Subnet mask 255.255.255.0
IP address
192.168.60.50
Subnet mask 255.255.255.0
Subnet mask 255.255.255.0
Subnet mask 255.255.255.0
Selecting duplicate IP addresses may cause conflicts on the network.
TCC-100 Operating Instructions