Thermo Scientific Dionex TCC-100 Operating Instructions Manual page 43

Thermostatted hplc column compartment
Table of Contents

Advertisement

The Address column lists all available modules. The entry that is available via Ethernet
(TCP/IP) includes the module's IP address. The entry for a module that is available via
USB (identified by the prefix "USB") includes the module's Ethernet address or serial
number.
If a module listed here is already connected to a Chromeleon server, the server name appears
in the Server Name column. If a module is available for connection, the column shows
0.0.0.0 or is empty.
The S/N column lists the module's serial number and the name of the computer to which the
module is connected while the Chromeleon server is running. The Ethernet column shows the
module's Ethernet address.
Select the desired column compartment by double-clicking. The module's address is
automatically entered in the Module Address field on the General tab page.
• Click the Firmware Download button to download the current firmware version from the
data system to the column compartment. The current firmware version is displayed here, as
well as the version of the file that is available in the \Bin directory of your Chromeleon
installation (TCC100.HEX).
Click OK to start the firmware download. Please note that the Chromeleon server needs to be
in running idle mode before the download can be started. The download takes several
minutes. To make sure that the firmware is downloaded correctly, do not perform further
actions while the download is running.
Important: During the download, make sure that communication between the
column compartment and Chromeleon is not interrupted and do not turn
off the instrument. Otherwise, the entire firmware may be lost. If the
download is not successful, contact Dionex Service.
TCC-100 Operating Instructions
37