TCC-100 Operating Instructions
• If necessary, copy the CmIPUtil program from the software CD to a separate directory on
your computer.
• Double-click the Setup.exe file to start installation. Follow the installation instructions as they
appear on the screen.
• Open the program via Start → Programs → Chromeleon → IP Utility → CmIPUtil.
The Module Information window of the IP Address Assignment dialog box lists all
modules that are active on the network, together with their IP address, serial number (S/N),
server name (the name or IP address of the server with which the module exchanges data),
gateway address, subnet mask, and Ethernet address. The IP address for modules that have
not yet been configured is 0.0.0.0 or 255.255.255.255 (→ Fig. 33).
• Enter the IP address and the subnet mask for the modules to be configured. The values depend
on the settings for the adapter for the instrument LAN (→ page 30).
• Select the module to be configured by its serial number.
• In the Available IP Addresses window, select the IP address to be assigned to the column
compartment. If it is not listed, click Add/Remove to enter the desired address.
• Enter the required subnet mask. Select the subnet mask assigned to the Chromeleon server PC.
When selecting the IP address of the network adapter of the server PC, be sure that only the
last digit differs from the IP address of the column compartment. Otherwise, it may not be
possible to address the connected Summit HPLC devices.
Fig. 33: Assigning IP properties (CmIPUtil)
32