Thermo Scientific Dionex TCC-100 Operating Instructions Manual page 33

Thermostatted hplc column compartment
Table of Contents

Advertisement

TCC-100 Operating Instructions
If you wish to connect several TCP/IP-enabled Summit HPLC devices to the server PC, you
must use a hub (10 Mbit) to establish the connection to the instrument LAN. In this case, use
standard network cables instead of the crossover cable (One standard network cable, part no.
8906.2038, is provided in the accessories kit.)
Fig. 28: Hub connection
Separate Network Interface Card (NIC)
A network connection that is independent of the office LAN is called an instrument LAN. This
kind of connection between the column compartment and the Chromeleon server requires the
installation of a separate 10 Mbit network interface card (NIC, in the operating system = adapter)
in the Chromeleon server PC.
Install the card according to the manufacturer's instructions, via Start → Settings → Control
Panel → Network → Adapters → Add.
Fig. 29: Installing the separate network interface card
27