GO-5000M-PMCL-UV User Manual (Ver. 1.1)
Timed Mode
This mode captures image(s) with a preset exposure time by using the external trigger. An additional
setting determines if the trigger pulse can be accepted during the exposure period.
Primary settings to use this mode
Acquisition Mode = Single frame, Multi-frame or Continuous
l
Trigger Mode = ON
l
Exposure Mode = Timed
l
Trigger minimum interval (Trigger Overlap = Readout) (1X8–1Y, 8-bit, CL Clock=72.85 MHz)
Trigger Mode
Timed Exposure Mode
Trigger Mode ON
t1
10μ (Min.)
Readout Mode
Full
AOI Center 2/3
AOI Center 1/2
AOI Center 1/4
AOI Center 1/8
V Binning ON (Full)
t2
18μ
- 76 -
Time (Min. Trigger Period)
9435us + 8.01 μs
6281us + 8.01 μs
4740us + 8.01 μs
2393us + 8.01 μs
1219us + 8.01 μs
4740us+ 8.01 μs
6L to 7L
Main Functions
t3