Setting analog output ranges
1. "1 Quick Start" > "04 Analog outputs" main menu
The menu includes one navigation line for each of the five analog output ranges that can
be set:
2. Set the output current range of an analog output.
– Open the parameter field you wish to edit: Press <ENTER>.
– Authorize access with Standard PIN, if necessary.
– Set the analog output range according to the table below:
3. Repeat step 2 for all additional analog output current ranges.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Structure of the navigation line for each analog output range
Component name
Specified analog output
range [mA]
4 to 20 (NAMUR)
4 to 20
0 to 20
20 to 0
2)
20 to 4
2)
CTRL = Function check
1)
Inversion
2)
Measuring range limit
Normal operation [mA]
Low limit
High limit
3.8
20.5
2
21
0
21
0
21
2
21
Functions
6.1 [1] Quick Start
Analog output range
Fault/CTRL
[mA]
1)
Low limit
High limit
3
21.5
2
21
0
21
0
21
2
21
87