6.9 [2.10] Settings > Correction of cross-interferences
3. Select source for correction value: "Source of correction value" parameter field
The following sources can be set for the correction of cross-interference:
– "Device-internal"
– "MODBUS TCP"
– "Analog input" (with installed option module 2)
Depending on the set parameter value, there will be different operating steps:
178
NOTICE
Malfunction of the device with correction of cross-interference via an analog input
The following applies to prevent the device from malfunctioning: Make sure that the
correction gases in → menus "2.10.3" to "2.10.6" and "2.09.1.n" match. For additional
information, see also → Set analog inputs (Page 137).
"Device-internal"/"Analog input"
Select an internally measured correction
value. "Component" parameter field.
The selection list contains all existing com-
ponents of the device except the respective
current measured component.
Enter substitute value for the measured
interference gas concentration.
"Substitute value" parameter field.
If the measurement of the interference gas
concentration fails, the set substitute value
is used for the correction of cross-
interference.
The measured value status is set to "Uncer-
tain".
The unit of the substitute value corresponds
to that of the configured interference gas.
Source of correction value
Determine external analyzer which delivers
the correction value via MODBUS TCP.
"MODBUS node" parameter field
Select external measured correction value
for correction of cross-interference via
MODBUS TCP.
"Measured value" parameter field.
Enter substitute value for the externally
measured interference gas concentration.
If the measurement of the interference gas
concentration fails, the set substitute value
is used for the correction of cross-
interference. The measured value status is
set to "Uncertain".
The unit of the substitute value corresponds
to that of the configured interference gas.