Siemens SIPROCESS GA700 Operating Manual page 145

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 6- 19
Function assignment at function (pre-selection): "Ext. funct. check"
Function
Trigger
Function check 1 to
8
"Level-triggered": activation by negative signal edge, deactivation by positive signal edge. Active at signal level "Low".
1)
Table 6- 20
Function assignment at function (pre-selection): "Calibration"
Function
Trigger
Calibration 01 to 36
Control calibration
"Edge triggered": Activation by rising signal edge.
1)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Remark / effect
External function check 1 ... 8: A different device signals "Function check".
Negative/positive
edge
Remark / effect
Calibration 01 to 36
Start single calibration. The calibrations must be assigned in the calibration
pool.
Positive edge
1)
Control of a calibration started via a digital input (Calibration 01 to 36) / start
of the next calibration:
This causes the switch to the next intermediate steps during calibration.
Positive edge
1)
Execution of the next step in the sequence is then time related.
Time 1 after the start of a calibration:
There is calibration gas in the device, the calculation is being made, the
signal filter is active.
Time 2:
Time 3 (only with 2-point calibration):
There is calibration gas in the device, the calculation is being made, the
signal filter is active.
Time 4:
2-point calibration is completed.
1-point calibration is completed (zero gas or span gas calibration)
2-point calibration: Injection of span gas necessary
6.8 [2.09] Settings > Inputs/outputs
Functions
145