7.2
OXYMAT 7 message list
Note
Message behavior at a restart
After a restart, the messages with the messages numbers 22 to 32 remain in the message
list. All other previous messages are deleted.
Table 7- 7
OXYMAT 7 message list
Message
No.
Text
2
Maximum warming-up time ex-
ceeded
3
Temperature on measuring
chamber too high
4
Temperature on measuring
chamber too low
5
Temperature on heating cartridge
too high
6
Temperature on heating cartridge
too low
9
Temperature on probe too high
10
Temperature on probe too low
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Cause
Device is operated outside the
•
permitted ambient temperature
range
Errors in temperature control
•
Device is operated outside the
•
permitted ambient temperature
range
Errors in temperature control
•
Temperature sensor of measuring
•
chamber has short-circuit/is faulty
Heating controller faulty
•
Temperature sensor of heating
•
cartridge is not connected/is faulty
Heating controller faulty
•
Temperature sensor of heating
•
cartridge is not connected/is faulty
Heating controller faulty
•
Device is operated outside the
•
permitted ambient temperature
range
Temperature control measuring
•
head faulty
Device is operated outside the
•
permitted ambient temperature
range
Temperature control measuring
•
head faulty
Alarm, error, and system messages
7.2 OXYMAT 7 message list
Remedy
Check ambient temperature
•
Check if there are additional mes-
•
sages and follow the instructions.
Check ambient temperature
•
Replace analyzer module
•
Check ambient temperature
•
Replace analyzer module
•
Check ambient temperature
•
Replace analyzer module
•
Check ambient temperature
•
Replace analyzer module
•
Check ambient temperature
•
Replace analyzer module
•
Check ambient temperature, replace
analyzer module, if necessary
257