[3.02] Maintenance & Diagnostics > Messages To Be Acknowledged; Displaying Messages To Be Acknowledged - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.16 [3.02] Maintenance & Diagnostics > Messages to be acknowledged
6.16
[3.02] Maintenance & Diagnostics > Messages to be acknowledged
6.16.1

Displaying messages to be acknowledged

Procedure
Note
Distinction
SIPROCESS GA700 distinguishes between messages that require acknowledgment and
messages that do not require acknowledgment.
Messages not requiring acknowledgement are cleared if the problem causing them no longer
exists. Messages that require acknowledgment are identified by the symbol □ and are
revoked when the cause of the message is no longer present and an acknowledgment is
made.
Note
Device behavior after voltage off/on
After the power supply has been switched off and on, messages that require
acknowledgment are treated as follows:
• Analyzer module(s): Messages that have not been acknowledged but require
acknowledgement are still pending.
• Device: Messages that have not been acknowledged but require acknowledgement were
deleted.
For additional information, see also → OXYMAT 7 message list (Page 257).
1. Main menu > "3. Maintenance & Diagnostics" > "2 Messages to be acknowledged"
The menu contains the following entries:
2. → Acknowledge message (Page 207). All messages to be acknowledged are listed
unfiltered.
3. → Acknowledge all messages (Page 208). All messages requiring acknowledgment can
be acknowledged in total with this function.
206
Structure of the navigation lines
1.
Messages to be acknowledged
2.
Acknowledge all messages
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05