6.9.6
OXYMAT 7
6.9.6.1
Assigning correction of cross-interference with a constant
Procedure with OXYMAT 7
1. Component > Main menu > "2. Settings" > "10. Correction of cross-interference" > "2
Constant"
2. Enable/disable correction of cross-interference: "Correction of cross-interference"
parameter field.
Authorize access with corresponding PIN, if necessary.
3. Enter interference gas concentration already known: "Interference gas concentration"
parameter field.
4. Define the coefficient.
Note
Operator control options
Refer to the coefficients stored in the device-internal reference table or enter a coefficient
manually.
Using the coefficients of the reference table:
– Select the interference gas you want to correct: "Interference gas reference table"
– Select the corresponding gas.
Using manually specified coefficient:
– "Interference gas reference table" parameter field: Set "User-defined" value.
– Enter coefficient already known manually: "Coefficient" parameter field.
5. Only with activated process tag switchover: Change the process tag assignment.
– Check the displayed process tag assignment.
– "Process tag" > "Change" parameter field.
– Activate the corresponding process tags that are to be integrated into the correction of
6. Enabling/disabling all corrections of cross-interferences (Page 179).
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
parameter field.
The editing window includes all gases for which a correction coefficient is available in
the reference table.
The selected coefficient is displayed in the "Coefficient" display.
The displayed editing window includes all available process tags.
cross-interference. Deactivate process tags that are not used.
6.9 [2.10] Settings > Correction of cross-interferences
Functions
169