Operation
4.3 Display
①
Start of scale
②
End of scale
③
Measuring range
④
High alarm limit
Figure 4-8
Example: Bar graph for measured-value display with display of measuring range limits
If not more than two components are displayed, the device displays measuring range limits
including high and low warning and alarm limits as bar graphs. The current measured value
in the example is 50% of the measured value.
Measured value status
①
Numerical value with unit, no qualifier displayed ⇒ Measured value status: "Good"
②
Numerical value with qualifier "?" and unit ⇒ Measured value status "Uncertain"
③
No numerical value displayed any longer, qualifier "- - -" ⇒ Measured value status "Bad"
Figure 4-9
Example: Display of measured value status
If a measured value reaches the measured value status "Uncertain" or "Bad", a "?" is
displayed behind the measured value. Measured values that can no longer be displayed are
displayed as "- - -".
64
⑤
High warning limit
⑥
Current measured value
⑦
Low warning limit
⑧
Low alarm limit
Operating Manual, 06/2017, A5E31930478-05
Operating with the Local User Interface