Siemens SIPROCESS GA700 Operating Manual page 146

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.8 [2.09] Settings > Inputs/outputs
Table 6- 21
Function assignment at function (pre-selection): "AutoCal"
Function
Trigger
AutoCal1/
AutoCal2
AutoVal1/
AutoVal2
AutoCal1 Sync./
AutoCal2 Sync.
(Function cannot
currently be execut-
ed)
AutoVal1 Sync./
AutoVal2 Sync.
(Function cannot
currently be execut-
ed)
"Edge triggered": Activation by rising signal edge.
1)
Table 6- 22
Function assignment at function (pre-selection): "Measuring range"
Function
Trigger
Measuring range
1 to 4
1)
Autoranging
2)
Switch measuring range 1 to 4.
1)
Activate autoranging.
2)
"Level-triggered": Activation by rising signal edge, deactivation by falling signal edge. Active at signal level "High".
3)
146
Remark / effect
Start AutoCal 1 or AutoCal 2.
Positive edge
1)
Start AutoVal 1 or AutoVal 2.
Positive edge
1)
Start AutoCal n/AutoVal n in sync.
Starting an AutoCal/AutoVal that is used on several networked devices
(chain) synchronously. To do so, one sync input and one sync output must
Positive edge
1)
be assigned per device.
The sync. indicates the start/end of a gas purge. With a 2-point calibration,
two sync sequences determine one single AutoCal sequence step. If such a
2-point calibration does not apply to one device in the chain, assign a pas-
sive calibration for the corresponding steps.
Positive edge
1)
Remark / effect
If the measuring range is not controlled according to the parameter assign-
ment but by means of a digital input: the measuring rang signaled as active
by the digital input is used ("Measuring range 1 to 4" or autoranging).
Positive/negative
If no digital input is active, the measuring ranges of the parameter assign-
edge
3)
ment are used accordingly.
If several digital inputs are active, the following prioritization applies: Meas-
uring range 4 → measuring range 3 → measuring range 2 → measuring
Positive/negative
range 1 → autoranging.
edge
3)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents