Alarm, error, and system messages
7.2 OXYMAT 7 message list
Note
Message no. 150
Information on parameters that have caused configuration problems can be found in the
→ [3.01] Maintenance & Diagnostics > Current messages (Page 203) menu. The parameters
causing the problems as well as the parameters affected by them are listed so that you can
start the corresponding remedial measures.
When the text "Remote service" is displayed, the parameters that cause the problems and
those affected by them can only be set with the remote service tool. Contact the service
technician in charge.
Message
No.
Text
151
Configuration restored following
fault
153
Factory synchronization of the
electronics faulty
160 to
Main. interval 1 to 16:
175
192 to
Ext. fault 1 to 8:
199
200 to
Ext. main. alarm 1 to 8:
207
208 to
Ext. main. demand. 1 to 8:
215
216 to
Ext. main. alarm 1 to 8:
223
224
Limit 1 violated
225
Limit 2 violated
226
Gas warning preliminary alarm
227
Gas warning main alarm
228 to
Ext. measured value violates limit: External measured value outside
235
236
AutoCal 1 aborted
262
Cause
Content of parameter memory was
faulty but has been recovered
Synchronization values of analyzer
module electronics faulty and cannot
be restored
Maintenance interval has expired
External fault signaled by digital input User-specific action
External maintenance alarm signaled
by digital input
External maintenance demanded
signaled by digital input
External maintenance required sig-
naled by digital input
Measured value outside limit 1
Measured value outside limit 2
Gas warning (preliminary alarm)
Gas warning alarm (main alarm)
warning limit 1 to 8
AutoCal could not be executed due to
a fault.
Remedy
Call list of current messages, look
for or filter for message type
"Config. warn.".
Call message detail view. The
affected parameter is displayed
here. Set the value of this param-
eter within the valid input limits.
Replace analyzer module
User-specific action
Reset expiration time of mainte-
nance interval.
User-specific action
User-specific action
User-specific action
User-specific action
User-specific action
User-specific action
User-specific action
User-specific action
Check and remedy pending errors
in the list of current messages.
Then: Execute AutoCal manually.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05