[2.20.06] Reference Gas Pump Oxymat 7 - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

6.14.4

[2.20.06] Reference gas pump OXYMAT 7

Requirements
Note
Availability of the menu
This menu is displayed only for OXYMAT 7 analyzer modules of the low-pressure version.
● You use an external reference gas pump.
Procedure
1. Main menu > "2. Settings" > "20. Service > "06. Pump reference gas".
This menu is only used to activate/deactivate the reference gas pump. Its performance
cannot be controlled.
2. Activate/deactivate reference gas pump: "Pump" parameter field.
– Authorize access with Expert PIN, if necessary.
– If you select the setting "On", you can read the activation status of the optional
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
pressure switch in the "Pressure switch for reference gas" display field.
Functions
6.14 [2.20] Settings > Service
199