Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Function (pre-selection) "DI links"
Purpose/resulting settings
Switch linked digital inputs.
1. "Function" parameter box:
Authorize access with corresponding PIN, if necessary.
2. Select one of the displayed links.
Additional linking properties are:
Type of logical operation (AND/OR)
Combining inputs
Function (pre-selection) "Other functions"
Purpose/resulting settings
Set additional functions.
1. "Function" parameter box:
Authorize access with corresponding PIN, if necessary.
2. Select one of the offered functions:
Sync. (Function cannot currently be executed): Synchronization of Au-
toCal sequences between several devices.
Calibration: Total function check: signaling of "Calibration" operating
state at device level.
Calib. AM1/AM2: Signaling of "Calibration" operating state at module
level.
Pump on/off: Switch external sample gas pump on/off.
Funct. check: Signal function check at device level.
Funct. check AM1/AM2: Signal function check at module level.
Signaling contact: Signaling contact: AutoCal signal contact, pulse
1000 ms.
6.8 [2.09] Settings > Inputs/outputs
Functions
159