Non-permissibility of calibrations

Overview
The tables below provide an overview of cases in which the execution of a calibration is not
permitted. Information on additional messages and the corresponding remedial measures is
available at:
● → OXYMAT 7 message list (Page 257)
● → ULTRAMAT 7 message list (Page 264)
● → CALOMAT 7 message list (Page 269)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Permissibility of calibrations with the following calibration drift messages
Text
Calibration tolerance exceeded at zero point.
Validation tolerance exceeded at zero point.
Calibration tolerance exceeded at span point.
Validation tolerance exceeded at span point.
Zero drift MR1 to MR4 outside tolerance.
Permissibility of calibrations with the following messages for measured values
Text
Measured value greater than full-scale value.
Cross-interference measured value uncertain or faulty.
Permissibility of calibrations with the following message for phase calibration or normali-
zation calibration
Text
Signal frequency changed. Normalization required!
6.7 [2.08] Settings > Calibrations
Functions
119