Siemens SIPROCESS GA700 Operating Manual page 267

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Mes-
sage
No.
Text
75
Temperature sensor on
AM PCB faulty.
80
Source current faulty.
81
Chopper faulty.
82
Reference voltage outside
tolerance.
83
Signal acquisition faulty
84
Test signal outside toler-
ance.
85
System error
86
System error
87
Parameter memory faulty. •
88
Analog output faulty.
89
ADC faulty.
95
Measured signal transient
reaction active.
128
SPI communication faulty. •
129
Communication with ADC
faulty
144
System error
145
Watchdog reset
146
Configuration faulty.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Cause
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Change the signal frequency
Turn the source off/on
Restart of chopper
Electronics of the analyzer module or
the option module defective
Analyzer module electronics defective •
Internal software error
If the error is signaled by the AM:
Analyzer module electronics defective
Internal software error (analyzer
module or processing module)
Content of parameter memory faulty
and cannot be restored
Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Remedy
Replace the analyzer module printed
circuit board or the analyzer module.
Check the status of the source in
"[3.05.1.1] Measured signal".
If necessary, replace the analyzer
module.
Replace the analyzer module.
Replace the analyzer module.
Replace the analyzer module.
Replace the analyzer module.
Restart the device.
Restart the device.
Replace the module.
Check the electrical circuit in the
menu "[3.11.1.1] Test analog out-
puts".
If necessary, replace the analyzer
module.
Replace the module.
Parameters of signal processing were
changed.
The measured value is invalid during
the subsequent settling time.
Replace the module.
Replace the analyzer module
Restart the device
Acknowledge the message.
Contact Service.
The active device configuration is
faulty.
Load the user or factory data in the
work data.
267