Mes-
sage
No.
Text
75
Temperature sensor on
AM PCB faulty.
80
Source current faulty.
81
Chopper faulty.
82
Reference voltage outside
tolerance.
83
Signal acquisition faulty
84
Test signal outside toler-
ance.
85
System error
86
System error
87
Parameter memory faulty. •
88
Analog output faulty.
89
ADC faulty.
95
Measured signal transient
reaction active.
128
SPI communication faulty. •
129
Communication with ADC
faulty
144
System error
145
Watchdog reset
146
Configuration faulty.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Cause
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
Electronic fault
•
Electronic fault
•
Change the signal frequency
•
Turn the source off/on
•
Restart of chopper
•
Electronics of the analyzer module or
the option module defective
Analyzer module electronics defective •
•
Internal software error
•
If the error is signaled by the AM:
•
Analyzer module electronics defective
Internal software error (analyzer
•
module or processing module)
Content of parameter memory faulty
•
and cannot be restored
Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Remedy
Replace the analyzer module printed
•
circuit board or the analyzer module.
Check the status of the source in
•
"[3.05.1.1] Measured signal".
If necessary, replace the analyzer
•
module.
Replace the analyzer module.
•
Replace the analyzer module.
•
Replace the analyzer module.
•
Replace the analyzer module.
•
Restart the device.
•
Restart the device.
•
Replace the module.
•
Check the electrical circuit in the
•
menu "[3.11.1.1] Test analog out-
puts".
If necessary, replace the analyzer
•
module.
Replace the module.
•
Parameters of signal processing were
•
changed.
The measured value is invalid during
•
the subsequent settling time.
Replace the module.
•
Replace the analyzer module
Restart the device
•
Acknowledge the message.
•
Contact Service.
•
The active device configuration is
•
faulty.
Load the user or factory data in the
•
work data.
267