[2.20.04] Calibrate Pressure Sensor; Overview - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.14 [2.20] Settings > Service
Note
Measured value errors
If you change the setpoints but do not normalize the module, the changed values are saved
and the measured value is calculated incorrectly.
After you have changed one or both setpoint settings, always run a normalization!
1. Main menu > "2. Settings" > "20. Service > "03. Normalization calibration".
The current values for "Sraw min" and "Sraw max" are displayed in the corresponding
display fields.
2. Enter setpoints:
– Authorize access with Expert PIN, if necessary.
– Enter setpoints: Use the parameter fields "Setpoint zero value" and "Setpoint end
3. Start normalization calibration wizard: "Start normalization" parameter field.
The newly determined values for "Sraw min" and "Sraw max" are displayed at the end of
the wizard process.
6.14.3

[2.20.04] Calibrate pressure sensor

6.14.3.1

Overview

This function calibrates the pressure sensor. Your device is then capable of correcting
fluctuations of the sample gas pressure.
In contrast to OXYMAT 7, ULTRAMAT 7 and CALOMAT 7 cannot be fitted with reference
gas flow monitoring. The reference gas flow monitoring is available as an option for
OXYMAT 7 analyzer modules. The information content of the module-specific service menus
differ accordingly.
Table 6- 49
Displayed parameter values
Actual value p sample gas
Actual value p2 reference gas
Start pressure sensor calibration
Offset p2 reference gas
Offset p sample gas
Linear factor p sample gas
Display of the reference gas values only with optional flow monitoring
1)
Button to start the calibration wizard
2)
Display of the menu only when equipped with optional pressure sensor
3)
196
value".
Display for calibration of the pressure sensor
OXYMAT 7
ULTRAMAT 7
Yes
Yes
Yes
No
1)
Yes
Yes
2)
Yes
No
Yes
Yes
No
Yes
Operating Manual, 06/2017, A5E31930478-05
CALOMAT 7
Yes
No
Yes
2)
2)
No
Yes
Yes
Operating with the Local User Interface
3)