Siemens SIPROCESS GA700 Operating Manual page 109

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

1. Component > Main menu > "2. Settings" > "04 Limits"
The menu has two navigation lines for setting alarm limits and/or warning limits. Each
limit range has a minimum and maximum limit that can be activated separately:
1)
2)
3)
2. Open selected limit range: Press <ENTER>.
3. Set values:
– Enable/disable values: "Min active" and/or "Max active" parameter fields.
– Authorize access with Standard PIN, if necessary.
– Enter values: "Minimum value" and/or "Maximum value" parameter fields
4. Set the monitoring range: "Limits violated at" parameter field.
Select a range in which there is a limit violation (limit violation caused by measured
values outside or within a band): → Table 6-2 Monitoring options using limits (Page 108).
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Structure of the navigation line for each limit
Limit number.
Message level
1)
Limit triggers an alarm or warning.
Monitoring indicator: Activation status of the minimum or maximum value or monitoring range
(monitoring inside/outside). Display in measuring mode only!
Display of minimum/maximum value with activation only, otherwise "-"
<
Minimum
<
>
value
>
2)
3)
2)
Functions
6.5 [2.04] Settings > Limits
Maximum
Unit
value
3)
Edit
109