CALOMAT 7 message list

Message
No.
Text
2
Maximum warming-up
time exceeded
11
Maximum temperature
exceeded. Heaters turned
off.
19
Temperature on meas-
urement block too high
21
Temperature on meas-
urement block too low
22
Calibration tolerance
exceeded at zero point
23
Validation tolerance ex-
ceeded at zero point
24
Calibration tolerance
exceeded at span point
25
Validation tolerance ex-
ceeded at span point
26
Measured value outside
the physical specifications
28
Control reserve used up
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Cause
Device is operated outside the permit-
ted ambient temperature range
Errors in temperature control
Safety function
Ambient temperature too high
One or more heater control loops
faulty
Device is operated outside the permit-
ted ambient temperature range
Error in the temperature control
Device is operated outside the permit-
ted ambient temperature range
Errors in temperature control
Zero gas changed, setpoint faulty
Drift heat conductor sensor
Calibration successful
Span gas changed, setpoint faulty
Drift heat conductor sensor
Calibration successful
Wrong gas mixture (extraneous gas-
es)
Wrong sample gas pressure:
pressure is outside the specification
High fluctuations in the thermal con-
ductivity sensor due to influence of
foreign matter, e.g. condensation res-
idues, dirt particles
Alarm, error, and system messages
7.4 CALOMAT 7 message list
Remedy
Check the ambient temperature of the
analyzer
Check if there are additional messag-
es and follow the instructions
Wait until the device has cooled down
Then acknowledge the message
You should also note the messages in
the logbook
Check the ambient temperature of the
analyzer
In the menu "[3.05.1.3] Heating con-
troller", check the temperatures of the
measurement block and analyzer
module
Check the ambient temperature of the
analyzer
In the menu "[3.05.1.3] Heating con-
troller", check the temperatures of the
measurement block and analyzer
module
Check the setpoint for the zero gas
calibration in the menu "[2.08.1.1]
Setpoints"
Carry out a zero gas calibration
Check the setpoint for the span gas
calibration in the menu "[2.08.1.1]
Setpoints"
Carry out a span gas calibration.
Check the sample gas concentration
and the sample gas pressure
Carry out a zero point and full-scale
value calibration if necessary
Check the zero gas and reference
gas supplies