Table 2-4
Table
0
1
2
[a] DEFAULT is a sensor calibration table where the reference calibration factor and calibration factors are 100%. This
There are also 17 sensor calibration tables named CUSTOM_0 through
CUSTOM_16. These tables do not contain any data when the N432A is shipped
from the factory.
Editing the sensor calibration tables require the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and calibration factor pairs
4 Save the table
Procedure
First, select the table you want to edit as follows:
1 Press
Figure 2-41
88
Installed thermistor sensor models
Sensor model
[a]
DEFAULT
478A
8478B
sensor calibration table can be used during the performance testing of the N432A.
>
Tables
Sensor Tbls screen
>
to display the Sensor Tbls screen.
Sensor Cal Tables
Keysight N432A User's Guide