Installing The Fixture On The Press Down Model (U9401A) - Agilent Technologies Medalist i1000 Installation Manual

In-circuit test system
Table of Contents

Advertisement

Installing the Fixture on the Press Down Model (U9401A)

1
Figure 24 Loosen alignment bars
2
3
4
5
Figure 25 View of honeycomb plate from the left
20
Loosen two of the alignment bars on the base of the Press Down Unit.
Disable the sensors by flipping the sensor switch down. (The sensor
switch is located at the rear of the Press Down Unit. See (a) in
Figure
43.)
When the sensors are disabled, the RETEST button blinks.
Place the fixture on the base of the Press Down Unit.
Press both DOWN buttons to lower the honeycomb plate towards the
fixture, but allow room for moving the fixture.
The honeycomb plate of the Press Down Unit has three holes along its
left and right sides
(Figure
the left and right sides of the top plate line up with the center holes of
the honeycomb plate.
25). Move the fixture so that the slots along
Medalist i1000 Installation Guide