Agilent Technologies Medalist i1000 Installation Manual page 28

In-circuit test system
Table of Contents

Advertisement

Releasing the Fixture
To release the fixture manually:
1
2
3
4
5
Figure 33
module holder
26
Press the MANUAL button to initiate manual mode.
Press MODUAL RELEASE to move the module holder to the unlocked
position
(Figure 34
on page 27).
Press and hold the MOTOR RELEASE button to disengage the module from
the tester interface.
Press MODUAL UP/DN to raise the module support, which moves the
module to the disengaged position
Press MODUAL RELEASE to move the module holder to the locked position
to hold the module in place.
fixture lock
module
support
(Figure 35
on page 27).
Medalist i1000 Installation Guide