Contents Product Identification Validity Intended Use Functional Principle Safety 1.1 Symbols Used 1.2 Personnel Qualifications 1.3 General Safety Instructions 1.4 Liability and Warranty Technical Data Installation 3.1 Vacuum Connection 3.2 Power Connection Operation Deinstallation Maintenance, Troubleshooting Returning the Product Disposal Literature For cross-references within this document, the symbol (...
Compact Cold Cathode Gauge IMG-500 Product Identification In all communications with Agilent, please specify the infor- mation given on the product nameplate. For convenient refer- ence copy that information into the space provided below. Agilent Technologies Model: Made in Liechtenstein...
Compact Cold Cathode Gauge IMG-500 Intended Use The Compact Cold Cathode Gauge IMG-500 has been designed for vacuum measurement of gases in the pressure range of … 1×10 2×10 mbar. The gauges can be operated in connection with an Agilent Vacuum Gauge Controller AGC-100, an Agilent Turbo AG Rack Controller, or with another controller.
Compact Cold Cathode Gauge IMG-500 Safety Symbols Used DANGER Information on preventing any kind of physical injury. WARNING Information on preventing extensive equipment and environ- mental damage. Caution Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage.
Compact Cold Cathode Gauge IMG-500 General Safety Instructions Adhere to the applicable regulations and take the necessary precautions for the process media used. Consider possible reactions with the product materials. Consider possible reactions (e.g. explosion) of the process media due to the heat generated by the product.
Compact Cold Cathode Gauge IMG-500 Liability and Warranty Agilent assumes no liability and the warranty becomes null and void if the end-user or third parties disregard the information in this document use the product in a non-conforming manner ...
Compact Cold Cathode Gauge IMG-500 Technical Data … 1×10 Measuring range (air, N 2×10 mbar ±30% (in the range Accuracy … 1×10 1×10 mbar) ±5% (in the range Reproducibility … 1×10 1×10 mbar) Output signal (measuring signal) 0 … +10.5 V Voltage range Voltage vs.
Page 12
Compact Cold Cathode Gauge IMG-500 ≤2 W Power consumption ≤1 AT Fuse 16 … 30 VDC Voltage at the supply unit with ≤ maximum cable length (ripple Adjustment the gauge is adjusted at the factory and requires no maintenance. Electrical connection...
Page 13
Compact Cold Cathode Gauge IMG-500 Temperatures +5 … +55 °C Operation Bakeout 150 °C (without electronics and magnetic shielding) -40 … +65 °C Storage ≤80% at temperatures Relative humidity ≤+31°C decreasing to 50% at +40°C Mounting orientation indoors only, altitude up to...
Page 14
Compact Cold Cathode Gauge IMG-500 Measuring Signal vs. Pressure Pressure p 1E+00 1E–01 mbar 1E–02 torr 1E–03 1E–04 1E–05 1E–06 1E–07 1E–08 1E–09 1E–10 0.0 0.5 1.0 1.5 10.0 10.5 Measurement signal U [V] p = 10 U = c + log...
Page 15
Compact Cold Cathode Gauge IMG-500 Gas Type Dependence Indicated pressure (gauge calibrated for air) p [mbar] Luft / Air Xe Kr Ne He –3 –4 –5 –6 –7 –7 –6 –5 –4 –3 –2 [mbar] In the range below 10 mbar, the pressure indication is linear.
Compact Cold Cathode Gauge IMG-500 Installation Vacuum Connection DANGER DANGER: overpressure in the vacuum system >1 bar Injury caused by released parts and harm caused by escaping process gases can result if clamps are opened while the vacuum system is pressurized.
Page 17
Compact Cold Cathode Gauge IMG-500 DANGER DANGER: protective ground Products that are not correctly connected to ground can be extremely hazardous in the event of a fault. Electrically connect the gauge to the grounded vacuum chamber. This connection must conform to...
Page 18
Compact Cold Cathode Gauge IMG-500 WARNING WARNING: electric arcing Helium may cause electric arcing with detrimental effects on the electronics of the product. Before performing any tightness tests put the pro- duct out of operation and remove the electronics unit.
Page 19
Compact Cold Cathode Gauge IMG-500 When making a CF flange connection, it can be ad- vantageous to temporarily remove the electronics and the magnet unit. Hex. socket AF 1.5 AF 7 Protective lid Keep the protective lid.
Page 20
Compact Cold Cathode Gauge IMG-500 Remove the protective lid and connect the product to the vacuum system. Seal with centering ring Protective lid Clamp Seal with centering ring and filter Keep the protective lid.
Compact Cold Cathode Gauge IMG-500 Power Connection Make sure the vacuum connection is properly made ( "Vacuum Connection"). If no sensor cable is available, make one according to the following diagram. Signal – 10 k Ident – Electrical connection Supply ( 15 …...
Compact Cold Cathode Gauge IMG-500 Operation When the supply voltage is applied, the measuring signal is available between pins 3 and 5. Over the whole measurement range, the measuring signal is output as a logarithm of the pressure (measuring signal vs. pressure "Technical Data").
Page 23
Compact Cold Cathode Gauge IMG-500 Ignition Delay When cold cathode measurement systems are activated upon switching the gauge on, an ignition delay occurs, which is typi- cally: 0.1 minute mbar 1 minute mbar 5 minutes 2×10 mbar Contamination Gauge failures due to contamination or wear and tear, as well as expendable parts (e.g.
Compact Cold Cathode Gauge IMG-500 Deinstallation DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before beginning to work, find out whether any parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.
Compact Cold Cathode Gauge IMG-500 Remove the gauge from the vacuum system and place the protective lid. When deinstalling a CF flange connection, it can be ad- vantageous to temporarily remove the electronics and the magnet unit ( 19).
Compact Cold Cathode Gauge IMG-500 Returning the Product WARNING WARNING: forwarding contaminated products Contaminated products (e.g. radioactive, toxic, caustic or microbiological hazard) can be detrimen- tal to health and environment. Products returned to Agilent should preferably be free of harmful substances. Adhere to the for-...
Compact Cold Cathode Gauge IMG-500 Disposal DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before beginning to work, find out whether any parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.