Agilent Technologies Medalist i1000 Installation Manual page 24

In-circuit test system
Table of Contents

Advertisement

Align the Sensors on the Press Down Model (U9401A)
After installing the fixture, follow these steps to align the two safety
sensors at the front of the Press Down Unit.
N O T E
1
2
3
Figure 28 Screws on sensor arm
4
5
6
7
22
Make sure the Medalist i1000 software is not running.
Disable the sensors by flipping the sensor switch down (see (a) in
Figure
43).
When the sensors are disabled, the RETEST button blinks.
Press both DOWN buttons to lower the press.
Loosen the screws on one of the sensor arms and adjust the position of
the sensor so that it is at the same height as the fixture support plate,
where the operator's hands would be when loading a board.
screws
Tighten the screws to secure the sensor in place.
Press the UP button to raise the press.
Reactivate the sensors by flipping the sensor switch up.
Check the RETEST button.
If it is blinking, the sensors are not aligned. Continue with
align them.
If it is not blinking, the sensors are properly aligned. Tighten the
screws on the other sensor arm if necessary, to secure the sensor in
place. This completes the alignment process.
sensor
Medalist i1000 Installation Guide
step 8